Section 8-5 - Covers
8-5-14
EMI Cover R (FRU No. 313)
Purpose: This is a description on how to remove and replace the EMI Cover R.
8-5-14-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-5-14-2
Needed Manpower
•
1person, 15 m travel
8-5-14-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-5-14-4
Removal Procedure
1.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43
2.) Unscrew seven (7) screws (1-7).
3.) Remove the EMI Cover R. Refer to Figure 8-62.
Figure 8-62 Removing the EMI Cover R
2
1
4
5
6
7
3