5-44
Section 5-10 - Service Platform
5-10-4-1
Logs
The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe,
Board, and DICOM.
NOTE:
Figure 5-34 provides a graphical example of the log entries for the System Logs.
Log table headings for the different logs are as follows:
•
System
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
•
Power
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
•
Infomatics
Log entry headings include TimeStamp, Revision, PtID, PtDOB, PtSex, PtWeight, PtHeight,
ExamID, Exam Category, ExamCurDate, and ExamStartTime.
•
Temperature
Log entry headings include Time Stamp; Error Level; Package; Upper FEC Sensor; and Lower FEC
Sensor.
•
Probe
Log entry headings include Time Stamp; Error Level; Package; Error Message; Severity; Revision;
and three (3) new labels that have not yet been named.
•
Board
Log entry headings include Time Stamp; Error Level; Package; Board; Severity; and two (2) new
labels that have yet been named.
•
DICOM
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...