Chapter 4 Functional Checks
4 - 57
4-13-14
PC2IP Assy function check procedure.
1.) Turn on the system and check the completion of Echoloader loading.
2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-42.
3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'PC2IP Full Test'
4.) Execute the diagnostic test.
All diagnostic test item of PC2IP should be passed.
5.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4-
7. Refer to Section 4-3-5 - System CFM and PWD Checks on page 4-8.
Figure 4-74 PC2IP Diagnostic test