Chapter 8 Replacement Procedures
8 - 37
8-4-11
Manu Key Assy (FRU No. 210)
Purpose: This is a description on how to remove and replace Manu key panel Assy.
8-4-11-1
Tools
•
Common pillips screwdrivers.
8-4-11-2
Needed Manpower
•
1person, 10 m travel
8-4-11-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
8-4-11-4
Removal Procedure
1.) Unscrew 12screws (1-12) from the manu key prefix bracket to remove it. Refer to Figure 8-39.
Figure 8-39 Unscrew 12 screws from the Menu key brkt
1
2
3
4
5
6
7
8
9
1 0
1 1
1 2
1
2
3
4
5
6
7
8
9
1 0
1 1
1 2