GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 10
Section 8-3 - Monitor
8-3-4
Monitor S/W Assy (FRU No. 103)
Purpose: This is a description on how to remove and replace the Monitor S/W Assy.
8-3-4-1
Tools
•
Common pillips screwdrivers
8-3-4-2
Needed Manpower
•
1person, 10 m travel
8-3-4-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-3-4-4
Removal Procedure
1.) Remove the Monitor Cover Front. Refer to section 8-3-3 on page 8.
2.) Disconnect two connectors (1-2). Refer to Figure 8-10.
3.) Unscrew two screws (3-4) from the Monitor Cover Front. Refer to Figure 8-11.
Figure 8-10 Disconnecting 2 cables
Figure 8-11 Unscrew 2 screws
1
2
3
4
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...