Chapter 4 Functional Checks
4 - 33
4-11-8
OP Panel LCD Assy Validation
1.) Boot up the system.
2.) Follow up the process from step1 to step7 of Section 4-11-1.
3.) Touch the touchpanel.
Figure 4-47 Touchpanel
NOTICE
In case of touch point is displayed in log of scfpapitest.exe while you press the touchpanel, LCD assy
is connected correctly.