8 - 12
Section 8-3 - Monitor
8-3-5
Speaker Assy (FRU No. 104)
Purpose: This is a description on how to remove and replace the Speaker Assy.
8-3-5-1
Tools
•
Common pillips screwdrivers
8-3-5-2
Needed Manpower
•
1person, 10 m travel
8-3-5-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Monitor Cover Set should be removed before proceed. For more information Refer to section 8-3-
8-3-5-4
Removal Procedure
1.) Disconnect two connectors (1-2) from the rear side of the Monitor. Refer to Figure 8-13.
Figure 8-13 Disconnect two connectors
1
2