Chapter 8 Replacement Procedures
8 - 199
8-11-3-3 Functional Checks for Service Platform (Diagnostic Program)
3.) The Service Login window for Service Platform will be shown on the monitor display.
4.) Select
GE Service
at the “Select User Level” field.
5.) Enter the password for the Service Platform.
6.) Click on
Okay
.
7.) Verify that the following screen (Service Platform) is displayed on the monitor.
Figure 8-241 Service Login
CAUTION
If the Login window for Service Platform is not displayed on the monitor, the installation of the
Service Platform has failed. Reload the application software or Base System Software (OS) +
application software. Contact a Technical Support for details.
Figure 8-242 GE Service
NOTICE
When the service platform is NOT displayed, check if
CAPS
lock is selected. The
CAPS
should not be
selected.