5.) Select the "dbeng7.exe" in the process list and click the "End Process" button.
6.) click the "Yes" button when "task manager warning" dialog box .
7.) Check if windows identify the both HDD.
8.) Run the window explorer. Move the cursor to bottom of the screen and click the "window explorer"
on the task bar.
9.) There should be 8 logical disk totally. Each HDD have 4 partitions, "SYSTEM", "USER",
"ARCHIVE", "SWAP". The used HDD which is set to slave is "H" to "K". And the new HDD in BEP
which is set to master is "C" to "F". (If the L5 system have MO drive, the slave drive number can be
"I" to "L")
10.)- The patient data files are in the "Idunn" directory in the "ARCHIVE" disk.
Figure 8-167 Task Manager with process
Figure 8-168 Dual Hard disk status
dbeng7.exe
End Process
Process
dbeng7.exe
End Process
Process