LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 56
Section 8-5 - Covers
8-5-9
OP Side R Cover (FRU No. 308)
Purpose: This is a description on how to remove and replace the OP Side R Cover.
8-5-9-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-5-9-2
Needed Manpower
•
1person, 25 m travel
8-5-9-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-5-9-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Front Base Cover (FRU No.306). Refer to 1.) on page 8-40.
4.) Remove the Front Cover (FRU No.306). Refer to section 8-5-7 on page 53.
5.) Remove the Keyboard Assy (FRU No,200). Refer to section 8-4-1 on page 16.
6.) Unscrew five (5) screws (1-5) and remove the Probe Holder Bracket. Figure 8-54.
Figure 8-54 Removing the Probe Holder Bracket
(1)
(2)
(3)
(4)(5)
Probe Holder Bracket