LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
3 - 10
Section 3-4 - Preparing for Installation
3-4-2
Physical Inspection
3-4-2-1
System Voltage Settings
Verify that the scanner is set to the correct voltage.
The Voltage settings for the LOGIQ™ 5 PRO Scanner is found on a label to the right of the Power switch
and External I/O, on the rear of the system.For the Rating Plate example, refer to
MARKINGS OF LOGIQ™ 5 PRO" on page 1-16
3-4-2-2
Video Formats
Check that the video format is set to the locally used video standard, NTSC or PAL.
3-4-3
EMI Protection
This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
WARNING
WARNING
Connecting a LOGIQ™ 5 PRO scanner to the wrong voltage level will most likely
destroy the scanner.