Chapter 4 Functional Checks
4 - 29
4-11-1 Main Keyboard Assy validation
6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the
scfpapitest program.
7.) Press the keyboard button and rotate rotary button. When you press or rotary keys, Log dialog box
displays the key log. Button name and displaying log name is like the table.
Figure 4-45 scfpapitest program
Button or
Rotary Name
The Name displayed in Log
Button or
Rotary Name
The Name displayed in Log
Record
External Output or VCR
button
B pause
B Pause Button
Userdefine 1
Userdefine 1 Button
Depth (Button)
Depth PushEncoder Button
Userdefine 2
Userdefine 2 Button
Depth (Rotary)
Depth PushEncoder Button
Userdefine 3
Userdefine 3 Button
B (Button)
B PushEncoder Button
3D
3D Button
B (Rotary)
B PushEncoder Rotary
Userdefine 4
Userdefine 4 Button
CF (Button)
CFM PushEncoder Button
Harmonic
Harmonics Button
CF (Rotary)
CFM PushEncoder Rotary
Table 4-25 Button and Rotary table