GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 145
4.) Disconnect all power connector from ATX smps in motherboard and hard disk and CD RW.
5.) Unscrew 2 screws on theATX power bracket.
6.) Take out the ATX smps.
7.) To remove Hard disk and CDRW Unplug the IDE cable for HDD or CDRW drive and power
connector.
8.) Unscrew the screws on the chassi for fixture.
9.) Take out CDRW or Hard disk.
Figure 8-159 Inside of BEP
Figure 8-160 Take out the ATX SMPS
Cable Tie
1
2
3
4
5
Power Bracket Screw
Cable Tie
1
2
3
4
5
Cable Tie
1
2
3
4
5
Power Bracket Screw
1
2
1
2
1
2
3
4
1
2
3
4
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...