GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207, R
EVISION
7
LOGIQ™5 PRO S
ERVICE
M
ANUAL
Chapter 5 Components and Functions (Theory)
5-5
Section 5-4
Front End
The Front End generates the strong burst transmitted by the probes as ultrasound into the body. It also
receives weak ultrasound echoes from blood cells and body structure, amplifies these signals and
converts them to an digital signal.
The digital representation of this signal is presented to the Mid Processor section.
•
RLY(Relay) : RLY Assy contains 3 connectors for probe interface and arrays of relays for switchable
connection. High voltage multiplexer provide connection between 128 signal of probe and
selectable 64 channels transmit/receive signal
•
LMT (Limiter) : This has transmit/receiving switch to isolate rx signal circuit from transmit pulse.
•
PMP(Preamplifier) : The preamplifier amplifies 64 echo signals. The amplified receiving signals are
sent to RDS.
•
ATD (Advanced Transmit Driver) :This has 64 channel Transmit pulse generator, bipolar pulse
drivers, delay controller.
•
RDS (Receiving Delay Summing) : Cascading two receiving beam formers (two RDSs) allows the
system to achieve 64ch delay summing.
•
FEC (Front End Control) : FEC control scan operation and transfer the scan parameters to others
board, ATD,RDS,RLY,PMP. Proper signal processing for tissue and dop are done in FEC
•
HV(High Voltage) : Generates THV and SHV for transmit pulse driver and High voltage mux.
•
SIG(Signal) : Receiving echo signal from PMP is transfer to RDS through this SIG
Figure 5-4 The Front End
Back plane Board
#6
F
EC
B
oa
rd
#7
HV Bo
ar
d
#5
R
D
S B
oar
d
#4
R
D
S B
oar
d
#3
CW
DOP Bo
ar
d
(O
pt
io
n)
#2
ATD Bo
ar
d
#1
PM
P B
oa
rd
LMT Board
RLY Bo
ard
SIG Board
Back plane Board
#6
F
EC
B
oa
rd
#7
HV Bo
ar
d
#5
R
D
S B
oar
d
#4
R
D
S B
oar
d
#3
CW
DOP Bo
ar
d
(O
pt
io
n)
#2
ATD Bo
ar
d
#1
PM
P B
oa
rd
LMT Board
RLY Bo
ard
SIG Board
DC Power Input
(3.3V, 6V,-6V,
15V,-15V,5V)
AC Power Input
(110V)
for HV
DC Power
output for
FAN(15v)
DC Power output for
Keyboard(15V,5V)
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...