8 - 140
Section 8-8 - PC Block
8-8-2-4-2
Removal of VIC board
1.) Disconnect all of connect in VIC assy.
2.) 5 connectors inside BEP. 14 pins connector to chassis from VIC. 50pin half pitch flat cable
connector. 50pin flat cable connector to PC2IP card. 2 pin connector to motherboard for power
control. 3 pin connector to motherboard for LED.
3.) Disconnect 2 connectors outside of BEP.
4.) Disconnect 2 connectors to motherboard.
Figure 8-150 Connectors on VIC assy
Figure 8-151 Video signal to Rear Panel & VGA to VIC
Figure 8-152 Disconnect 2 connectors
2309649
VG
A to VIC
J
2
Video signal to Rear panel
2309649
VG
A to VIC
J
2
Video signal to Rear panel
J8 ,
J12
J8 ,
J12