Chapter 8 Replacement Procedures
8 - 11
4.) Remove the Monitor S/W Assy. Refer to Figure 8-12.
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-3-4-5
Mounting procedure
Install the new parts in the reverse order of removal.
Figure 8-12 Removing Monitor S/W Assy
Table 8-5
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-3-4. Equipment
passed all required tests and is ready for use. “
CRT Assy function check procedure
3
4