Section 4-10 - Monitor Function Checks
Section 4-10Monitor Function Checks
4-10-1
CRT Assy function check procedure
1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.
4-10-2
Monitor Cable Assy function check procedure
1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.
4-10-3
Monitor S/W Assy function check procedure
1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.
Figure 4-40 Monitor Test pattern screen
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...