GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 8
Section 8-3 - Monitor
8-3-3
Monitor Cover Set (FRU No. 102)
Purpose: This is a description on how to remove the Monitor Cover Set.
8-3-3-1
Tools
•
Common pillips screwdrivers
8-3-3-2
Needed Manpower
•
1person, 10 m travel
8-3-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-3-3-4
Removal Procedure
1.) Remove the Tasklamp Assy, For more information, refer to section 8-3-6 on page 14.
2.) Unscrew three screws (1-2,6) to remove the Monitor Cover left.
3.) Unscrew three screws (3-5) to remove the Monitor Cover right.
4.) Unscrew four screws (7-10) to remove the Monitor Cover front.
5.) Monitor Cover Top can be removed without unscrewing.
Figure 8-9 MONITOR COVER SET DISASSEMBLY
7
8
9
10
1
2
3
4
6
5
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...