Chapter 5 Components and Functions (Theory)
5-37
5-8-4
AC Power Outlet for Peripheral
The REAR ASSY and REAR CONN 220V ASSY are the rear-panel assemblies which interconnect
external signals and power with the LOGIQ 5 system. Both of the assemblies are the same except the
circuit breaker’s current ratings depending on voltage of the system power source (See Safety
described below).
5-8-4-1
Safety
•
Circuit Breaker
•
Electrical Isolation
The rear-panel's Insite (modem) interface is electrically isolated from the system's internal circuit.
The isolator's dielectric strength should be 500VAC RMS or higher within 1 minute (in compliance
with IEC60601-1-1).
5-8-4-2
Camera (Printer) Remote Control Interface
The Rear-panel has two remote control interface terminals for B/W and color printers.
Table 5-11 Circuit Breaker Specification
Assembly
Power Voltage Rating
(V)
Peripheral Current
Rating (A)
Trip Time
Reset Time
REAR CONN
110
5.0
1 hour max. @ 5.4A, 25
degreeC
60 seconds
REAR CONN 220V
220
3.0
1 hour max. @ 3.5A, 25
degreeC
180 seconds
Table 5-12 Camera (Printer) Remote Control Interface Specifications
Level (V)
Active Polarity
Retention Time
Color camera (shutter) control
H:2.0/L:0.8
L
60 msec
B/W camera (printer) control
H:2.0/L:1.0
L
100 msec
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...