GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 60
Section 8-5 - Covers
8-5-11
OP Bottom Cover (FRU No. 310)
Purpose: This is a description on how to remove and replace the OP Bottom Cover.
8-5-11-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-5-11-2
Needed Manpower
•
1 persons, 25 m travel
8-5-11-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-2-2 on
page 5.
8-5-11-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Front Base Cover (FRU No.305). Refer to 1.) on page 8-40.
4.) Remove the Front Cover (FRU No.306). Refer to section 8-5-7 on page 53.
5.) Remove the OP Side L Cover (FRU No.307). Refer to section 8-5-8 on page 54.
6.) Remove the OP Side R Cover (FRU No.308). Refer to section 8-5-9 on page 56.
7.) Unscrew four (4) screws (1-4). Refer to Figure 8-58.
Figure 8-58 Removing the OP Bottom Cover
1
3
2
4