GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 33
8-4-9
OP Panel Trackball Assy (FRU No. 208)
Purpose: This is a description on how to remove and replace OP Panel Trackball Assy.
8-4-9-1
Tools
•
Common pillips screwdrivers.
8-4-9-2
Needed Manpower
•
1person, 5 m travel
8-4-9-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
8-4-9-4
Removal Procedure
1.) Disconnect two connectors (1-2) from the Trackball Assy. Refer to Figure 8-34.
2.) Unscrew three screws (3-5) to remove the Trackball Assy. Refer to Figure 8-35.
Figure 8-34 Disconnect two connectors
Figure 8-35 Unscrew 3 screws