GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 40
Section 8-4 - Keyboard Block
3.) Disconnect two cable connector. Unscrew ten(10) screws (1-10)
to remove the Menu Encoder
Assy
.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-12-5
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-42 Unscrew 4 screws from Menu Encoder Circuit Board
Table 8-19 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
Equipment passed all required tests and is
ready for use. “
Section 4-11-11
Menu Encoder Assy
1
2
3
4
5
6
7
8
3
10
1
2
3
4
5
6
7
8
3
10
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...