LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 152
Section 8-8 - PC Block
8-8-3-6-4
Check the patient data
1.) Turn on the system
2.) If system boot up, select patient menu and confirm that the all of patient data are available.
3.) The database utility should be enabled by the network IP address setting procedure. If this “save
setting procedure have been missed, the patient data image can not be show up since the database
link error
8-8-3-6-5
Shut down the system.
1.) Turn off the circuit breaker.
2.) Disconnect the slave HDD(used HDD) from HDD cable in BEP assy carefully.
3.) Close the BEP cover and all other cover.
4.) Turn on the system and boot up.
5.) Check the system operation.