Chapter 10 Care & Maintenance
10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Monthly Quarterly
Per Facilities
QA Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor and Touch Panel
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Control Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks
•
also after corrective
maintenance
Surface Probe Leakage Current Checks
•
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
•
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
•
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
•
also after corrective
maintenance
Measurement Accuracy Checks
•
also after corrective
maintenance
Probe/Phantom Checks
•
also after corrective
maintenance
Functional Checks
•
also after corrective
maintenance
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...