Chapter 8 Replacement Procedures
8 - 83
8-6-4
Rear Caster (FRU No. 404)
Purpose: This is a description on how to remove and replace the Rear Caster.
8-6-4-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
•
Lifter (180Kg)
8-6-4-2
Needed Manpower
•
1 person, 15 m travel
8-6-4-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Lift the system before proceed.
8-6-4-4
Removal Procedure
1.) Unscrew eight hexagonal bolts (1-8).
2.) Remove the Rear Caster. Refer to Figure 8-76.
Figure 8-76 Removing the Rear Caster
1
2
6
8
7
5
4
3