Chapter 4 Functional Checks
4 - 31
4-11-1 Main Keyboard Assy validation
8.) Check the 'Fast' radio button into Indicator box on scfpapitest program and press 'Set All' button.
Figure 4-46 Fast button
NOTICE
After pressing Fast radio button, indicators should be blanked on the frontpanel.