Chapter 4 Functional Checks
4 - 13
4-3-8
Using Cine
4-3-8-1
Activating CINE
Press
FREEZE
, then roll the
TRACKBALL
to activate CINE. To start CINE Loop playback, press the
left most button on the menu key panel of keyboard. To stop CINE Loop playback. press Run/Stop.
4-3-8-2
Quickly Move to Start/End Frame
Press FIRST to move to the first CINE frame; press LAST to move to the last CINE frame.
4-3-8-3
Start Frame/End Frame
Turn the
START FRAME
dial to the left to move to the beginning of the CINE Loop. Turn the dial to the
right to move forward through the CINE Loop.
Turn the
END FRAME
dial to the right to move to the end of the CINE Loop. Turn the dial to the left to
move backward through the CINE Loop.
4-3-8-4
Adjusting the CINE Loop Playback Speed
Turn the
LOOP SPEED
dial right/left to increase/decrease the CINE Loop playback speed.
4-3-8-5
Disconnecting B-Mode CINE from Timeline CINE
To review the B-Mode CINE Loop only, press CINE MODE SELECTION and select B ONLY.
To review the Timeline CINE Loop only, press CINE MODE SELECTION and select TL ONLY.
To return to linked B-Mode and Timeline CINE Loop review, press CINE MODE SELECTION and select
B/TL.
4-3-8-6
Moving through a CINE Loop Frame By Frame
Turn
FRAME BY FRAME
to move through CINE memory one frame at a time.
4-3-9
Image Management (QG)
For Image Management functionality refer to the LOGIQ™ 5 PRO Quick Guide. It talks about several
topics:
•
Clipboard
•
Printing Images
•
Browsing and Managing an Exam’s Stored Image
•
Connectivity, and Dataflow Concept and Creation
•
Starting an Exam
•
Configuring Connectivity
•
TCP/IP
•
Services (Destinations)
•
Buttons
•
Views
Verifying and Pinging a Device
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...