8 - 24
Section 8-4 - Keyboard Block
8-4-5
OP Panel Encoder Assy (FRU No. 204)
Purpose: This is a description on how to remove and replace OP Panel Encoder Assy.
8-4-5-1
Tools
•
Common pillips screwdrivers.
•
Longnose.
8-4-5-2
Needed Manpower
•
1persons, 5 m travel
8-4-5-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
8-4-5-4
Removal Procedure
1.) Remove the eight(8) encoder knobs (1-8) from the front side of the Keyboard. Refer to Figure 8-24.
Figure 8-24 Removing 8 encoder knobs
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Содержание LOGIQ 5 PRO