Chapter 8 Replacement Procedures
8 - 51
8-5-6
Front Base Cover (FRU No. 305)
Purpose: This is a description on how to remove and replace the Front Base Cover.
8-5-6-1
Tools
•
Common pillips screwdrivers
8-5-6-2
Needed Manpower
•
1person, 15 m travel
8-5-6-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-5-6-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Unscrew four screws (1-4). Refer to Figure 8-49.
Figure 8-49 Unscrew 2 screws on the Right side
(1)
(2)
(3), (4) are on the left side
Right side