Section 8-4 - FRU Replacement Procedures
8-4-3
DC Power Replacement Procedure
8-4-3-1
Manpower
One person, 30 minutes
8-4-3-2
Tools
4 mm blade screwdriver.
8-4-3-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove Right Side Cover as described in
"Side Covers Removal Procedure" on page 8-3
.
3.) Remove Card Rack Cover.
a.) Turn seven (7) quarter-turn screws to loosen Card Rack door.
b.) Holding the door, lift it out of its retainers along the lower edge of Card Rack.
8-4-3-4
DC Power Removal Procedure
1.) Unscrew three (3) screws as shown in
Figure 8-30
.
2.) Unplug PCI cable from the socket on IMP2 board.
3.) Pull the DC Power a little to get access to power cables.
4.) Remove retaining clasp and unplug Power Cable and HV Cable.
5.) Remove DC Power.
DANGER
Dangerous voltages are present in this equipment. Completely power off and
shut down the scanner before servicing.
Figure 8-30 DC Power and Cabling.
Screws
HV power cable
DC power cable
PCI cable
IMP2 Board
DC Power