Chapter 8 - Replacement Procedures
8 - 35
8-4-2-5
AC Power Installation Procedure
1.) Connect the cables. You may need to leave the Speed Control cable and the Fan cable
unconnected until you have fastened the unit. Input Voltage connector and socket are coded to
prevent mistakes. In 115V system the connector matches one end of the 8-holed socket. In 230V
system it matches the other end of the socket.
Figure 8-29
is on a 230V system, and the connector
can only be plugged in the right end of the socket.
2.) Fasten AC Power Module for correct Voltage and connect the remaining cables.
3.) Install Lower Rear Cover as described in
"Lower Rear Cover Installation Procedure" on page 8-8
.
4.) Install Filter Cover as described in
"Filter Cover Installation Procedure" on page 8-6
.
5.) Install Right Side Cover as described in
"Side Covers Installation Procedure" on page 8-4
.
6.) Do functional check-out as described in Chapter
4
..
DANGER
Before installing a replacement unit, verify that the power selector on the AC
Power Module is set to the correct voltage for the peripherals used on the
system.
Figure 8-29 Input Voltage connector/socket
4 pin connector
8 holed socket
Red blocks prevents wrong connections