GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5 - 69
LD5
- Yellow 2
DSP1 has received data
ON in Doppler mode, else OFF
LD6
- Yellow 3
Data Out on Pipelink
ON during data transfer to Pipelink
FIFO
LD7
- Yellow 4
Audio Path OK
ON in Doppler Mode and audio
present.
Lower Batch of LEDs
LD8
- Red
Board Failure
Normally OFF
ON if errors detected
LD9
- Green
Board OK
Normally ON
OFF if errors detected
LD10
- Yellow
DSP2 Local Software Running
Flashing
LD11
- Yellow
Audio Path OK
ON in Doppler Mode and audio
present.
LD12
- Red
Pipelink Clock Problem
OFF
ON if SDP clock is out of phase with
RFT clock.
NOTE: If system is working and LED is
ON: System OK.
Table 5-53 LEDs on the SDP Board
LED #
LED Color
Description
Normally State