Chapter 8 - Replacement Procedures
8 - 41
8-4-5-4
Rear Casters Removal Procedure
1.) Turn Front Casters as shown in
Figure 8-34
.
2.) Activate Direction Lock.
3.) Put the bevel-edged board on the floor.
4.) Pull the system backwards up the board incline. This will lift the Rear Wheel on the opposite side of
the System from the floor.
WARNING
WARNING
Prior to elevating scanner, verify that the floating keyboard is locked in its lowest
position.
WARNING
WARNING
Use extreme caution as long as VIVID 7 is un-stable, not resting on all four Casters.
Figure 8-34 Pull system backwards up the board incline