8 - 30
Section 8-3 - SW Loading Procedure
8-3-10-4 Restore Standard Printer IP and Port Settings (cont’d)
9.) Verify
Printer Name or IP Address
as recorded in
Table 8-8 on page 8-21
10.)Select
Raw
.
11.)Verify
Port Number
as recorded in
Table 8-8 on page 8-21
.
12.)Select OK.
13.)Close rest of the Windows by selecting OK.
8-3-10-5
DC Offset Alignment
1.) From the upper section of the window, select TEST.
2.) Then select GO TO SYSTEM TEST.
3.) From the Module selection block, select DC OFFS. CAL.
4.) Reboot the System.
8-3-10-6
Verify System Software Version
1.) From the alphanumeric keyboard on the Operator Panel, select
CONFIG
(F2 key), from the
operator log in window, select ADM
from the pull down menu, then select LOG ON
.
(No password
Needed).
2.) Select ABOUT tab.
3.) Verify that the SW versions is the same as what was recorded in
Table 8-10 on page 8-23
.
•
Torque wrench (available range: 12 Nm) with 5 mm hexagon bit
4.) Torque wrench with 5 mm hexagon bit to tighten the bolts to 12 Nm.
Figure 8-25 Configure Port
Raw
Port Number
OK