Chapter 3 - Installation
3 - 27
3-8-1-5
Units of Measure
Figure 3-19 Select Units of Measure
Table 3-15 Select Units of Measure
Step
Task
Expected Result(s)
1.
Open the Configuration Window as described in
3-8-1-1
"Log on
to the system" on page 3-22
.
The System Settings Window is displayed.
2.
Use the drop down dialog to select Metric or US Units.
The selected units (Metric or US) will be used for measurements as soon
as the unit has been restarted.
3.
Restart the scanner.
All the changed settings will be used after the restart.
Units