GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
8 - 18
Section 8-3 - SW Loading Procedure
8-3-9
Recording of SW Option Keys, TCP/IP and Service Settings
NOTE:
Service Key is not required and should NOT be installed prior to Upgrade
.
8-3-9-1
Record SW Option Keys
1.) Select
CONFIG
(F2) on the alphanumeric Keyboard on the operator panel.
2.) From the Operator Login window select
adm
from the Operator pull-down menu.
3.) Select LOG ON.
4.) From ADMIN tab, select the SYSTEM ADMIN.
5.) Record SW Options Keys from SW Options Key window.
Figure 8-1 System Admin
WARNING
WARNING
Make sure that the SW Option Keys (passwords) have been recorded correctly. If the
key is incorrect, you will not be able to log on after the SW installation has been
completed. The password is case sensitive. Hyphens must also be recorded. There may
be more than one password.
Table 8-6
SW Options Keys
SW OPTIONS KEYS