GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
7 - 60
Section 7-4 - Acquisition Diagnostics
7-4-3
Beamformer Calibration
When following the steps in
7-4-2 - Start System Test
on page
7-59
,
the default screen will be the
Calibration Screen.
From the Calibration Screen you can;
•
Do DC Offset Calibration (DC Offs. Cal) of the A/D converters in the Front End.
•
Reset the calibration (Cal. Reset) to the factory default setting.
•
Show a graph regarding the calibration (Show graph).
•
Deselect/Select channels are used in manufacturing to isolate noisy channels (Channel Control).
With a linear probe connected to the system, it is possible to disconnect groups of channels either
on the TX Board or on the RX Board to isolate the noise source.
•
The buttons in the lower part of the screen are used for board testing (see
7-4-4 - System Test
on
page
7-61
).
To start a test, click the respective button.
The “LEDs” in
Figure 7-55
will change color when a test has been done;
•
Green indicates that the test passed.
•
Red indicates that the test failed.
Figure 7-55 Calibration Dialog
Channel Control
(used for factory
tests)
DC Offs. Cal
Cal. Reset
Show graph
“LEDs”
Gray - Test not done
Green - Test Passed
Red - Test failed