3 - 2
-
3-2-2
Installation Warnings
DANGER
When using any test instrument that is capable of opening the ac ground line
(i.e., meter’s ground switch is open), don’t touch the unit!
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
its operating environment.
Table 3-3
Time for Settlement
°
C
60
55
50
45
40
35
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
-35
-40
°
F
140
131
122
113
104
96
86
77
68
59
50
41
32
23
14
5
-4
-13
-22
-31
-40
hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20
CAUTION
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
CAUTION
Do not wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
CAUTION
Do not operate this unit unless all board covers and frame panels are securely in place. System
performance and cooling require this.
CAUTION
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the Vivid 7 / Vivid 7
PRO and kept near the unit for quick reference.
CAUTION
Acoustic Output Hazard
Although the ultrasound energy transmitted from the Vivid 7 / Vivid 7 PRO probe is within AIUM/
NEMA standards, avoid unnecessary exposure. Ultrasound energy can produce heat and
mechanical damage.