Section 8-3 - SW Loading Procedure
8-3-10-3
Verify SW Key Option Settings
1.) From the alphanumeric keyboard on the Operator Panel, select
CONFIG
(F2 key), from the
operator log in window, select
adm
from the pull down menu, then select LOG ON
.
(No password
Needed).
2.) From the bottom of the configuration page select the CONNECTIVITY
tab.
3.) From the lower part of the window, select the SERVICE
tab.
.
4.) From Video Settings pull down menu, verify
Format
as recorded in
Table 8-9 on page 8-23
.
5.) From the VCR pull down menu, verify
VCR type
as recorded in
Table 8-9 on page 8-23
.
6.) From the top of the connectivity tab, select the TCPIP
tab.
Figure 8-20 Video Settings