Section 10-6 - Using a Phantom
10-7-9
Probe Leakage Current Test
10-7-9-1
Definition
This test measures the current that would flow to ground from any of the probes through a patient who
is being scanned and becomes grounded by touching some other grounded surface.
10-7-9-2
Generic Procedure
Measurements should be made with the ground open and closed, with power line polarity normal and
reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst
case condition.
NOTE:
Each probe will have some amount of leakage current, dependent on its design. Small variations in
probe leakage currents are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement.
Figure 10-5 Set Up for Probe Leakage Current
POWER
OUTLET
H (BLACK)
POLARITY REVERSING SWITCH
MOMENTARY
SWITCH
N (WHITE)
G (GREEN)
LEAKAGE TEST
METER
CONSOLE
PROBE