GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
2 - 10
Section 2-3 - Facility Needs
2-3-5
Networking Pre-installation Requirements
2-3-5-1
Stand Alone Scanner (without Network Connection)
None.
2-3-5-2
Scanner Connected to Hospital’s Network
Supported networks:
DICOM Network (option)
2-3-5-3
Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-5-4
DICOM Option Pre-installation Requirements
To configure the Vivid 7 / Vivid 7 PRO to work with other network connections, the site’s network
administrator must provide information to complete the form in
Figure 2-2 - Worksheet for DICOM
Network Information
. Ensure that there are no spaces in any field of the form.
Entries must include:
•
A host name, local port number, AE Title, IP address and Net Mask for the Vivid 7 / Vivid 7 PRO.
•
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
•
The host name, IP address, port and AE Title for each device the site wants connected to the Vivid
7 / Vivid 7 PRO
for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and
the revision of the device, is also included. This information may be useful for error solving.