Chapter 3 - Installation
3 - 13
3-7-9
Probe Connection
Vivid 7 / Vivid 7 PRO has four positions to plug in probes. The connector to the left is a “dummy”
connector, only used for parking one probe. The three other connectors are selectable and used for
scanning.
A separate “Pedof” Doppler probe connector is also available.
3-7-9-1
Connect a Probe
NOTE:
It is not necessary to turn OFF Power to connect or disconnect a probe.
1.) Hold the probe connector vertically with the cable pointing upward.
2.) Turn the connector locking handle counter-clockwise to the horizontal position.
3.) Align the connector with the probe port and carefully push into place.
4.) Rotate the locking handle to the full vertical position to lock in place. (See
Figure 3-10
.)
5.) Position the probe cable so that it is not resting on the floor.
3-7-9-2
Disconnect Probe
1.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
(See
Figure 3-10
.)
2.) Remove the connector from the port.
3.) Ensure that the probe head is clean before placing the probe in its storage case, see
section 10-5-
4 on page 10-7
for cleaning instructions.
Figure 3-10 Probe Connectors
CAUTION
Do not allow the probe head to hang freely. Excessive impact to the probe will result in
irreparable damage.
CAUTION
To prevent probe connector pins, or PCB board damage, do not use excessive force when
connecting the probes.
CAUTION
Take the following precautions with the probe cables:
Keep away from the wheels
Do not bend
Do not cross cables between probes.
“Pedof” Doppler Probe Connector
Probe Connectors
Parking Slot
for probe
connector
Locked position
Unlocked position