GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
8 - 34
Section 8-4 - FRU Replacement Procedures
8-4-2
AC Power Replacement Procedure
8-4-2-1
Manpower:
One person, 1 hour.
8-4-2-2
Tools:
Phillips screwdriver size 2
8-4-2-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove Right Side Cover as described in
"Side Covers Removal Procedure" on page 8-3
.
3.) Remove Filter Cover as described in
"Filter Cover Removal Procedure" on page 8-6
.
4.) Remove Lower Rear Cover as described in
"Lower Rear Cover Removal Procedure" on page 8-8
.
5.) Note Voltage type of AC Power Module being replaced (see label on rear of system).
8-4-2-4
AC Power Removal Procedure
1.) Disconnect Speed Control cable, Fan cable and Input Voltage cable.
2.) Remove four (4) screws holding AC Power.
3.) Turn AC Power Module to get better access and unplug the rest of the cables. Make sure each
connector and socket are properly marked, and easy to identify during installation.
DANGER
Dangerous voltages are present in this equipment. Completely power off and
shut down the scanner before servicing.
Figure 8-28 AC Power Module
Fan Cable
Speed Cable
Screws
Peripheral Voltage Select
Box turned for access to cabling.
Input Voltage cable