GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
5 - 46
Section 5-11 - Circuit Boards
5-11-2-3
Inputs
DC Voltages
Input Signals
Control Signals
5-11-2-4
Outputs
Output Signals
5-11-2-5
Fuses, Jumpers and Dip-switches
None
5-11-2-6
LEDs
None.
Table 5-23 DC Output Voltages, Plug P3 & P4
Input
Description
Connected from:
+ 5 Va
DC output voltage, distributed via Mother Board (Back
Plane).
The “a” indicates that this voltage is used for analog circuits
DC Power Supply > Backplane
- 5 Va
DC output voltage, distributed via Mother Board (Back
Plane).
The “a” indicates that this voltage is used for analog circuits.
DC Power Supply > Backplane
GND
Distributed via Mother Board (Back Plane)
DC Power Supply > Back Plane
Table 5-24 Input Signals
Input
Description
Connection from:
EL1-128
Received signals (echoes) and Transmitted pulses (1 of 3
probe connectors)
The active probe
Table 5-25 Input Signals
Input
Description
Connection from:
Front End Data Bus
Digital data bus used for control of the boards in the Front
End
RLY > Backplane > FEC
Probe MUX Control
Probe Configuration
RLY > Backplane > FEC
Table 5-26 Output Signals
Output
Description
Connection to:
XD1 - 128
128 signals to TX-128 and RX-128
XD Bus Boards
CW or PW TX
CW or PW Doppler transmitter pulses
Stand Alone Doppler probe
CW or PW RX
CW or PW Doppler RX signals (echoes)
Stand Alone Doppler probe