Check the Connection to the DICOM Server (Ping)
1.) After the system is rebooted, press Config, log on as adm (no password) and enter the Connectivity
screen.
2.) Click the VIEWS tab and the DATAFLOW button and you will see all available data flows.
3.) Click on the + by the “DICOM Server” dataflow to see the contents of that dataflow. This dataflow
has one DICOM Server with Storage Service.
Two new lines will appear, DICOMSERVER and the name of the server (e.g. ProSolv server).
..
4.) Click once on DICOMSERVER to highlight the line, then click on the Check button.
You should get a green check mark.
This is a hardware “ping” indicating that you have contact with the DICOM server’s hardware. If you
get a red cross, double check cabling and TCPIP addresses).
5.) Click once on the next line (e.g. ProSolv server) to highlight, then click on CHECK.
You should get a green check mark.
This is a DICOM “ping” indicating that you have contact with the DICOM server software. If you get
a red cross, double check AE Title and port number).
Figure 3-57 All Data flows - Select “DICOM Server”
Figure 3-58 Views and Dataflow Selections
VIEWS tab
DATAFLOW button
+
DICOM server
Storage service