GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 3 - Installation
3 - 9
3-7-2
Electrical Specifications
3-7-2-1
Verification of the System’s Voltage Settings
Verify that the mains voltage specified for the unit is available on site.
The Voltage settings for the unit is found on the label to the left of the Mains Power Circuit Breaker on
the rear of the system, see
Figure 3-3 on page 3-9
.
3-7-2-2
Electrical Specifications for Vivid 7 / Vivid 7 PRO.
WARNING
WARNING
CONNECTING A Vivid 7 / Vivid 7 PRO UNIT TO THE WRONG VOLTAGE LEVEL WILL
MOST LIKELY DESTROY THE UNIT.
Table 3-7
Electrical Specifications for Vivid 7 / Vivid 7 PRO
GE VINGMED
PART NUMBER
DESCRIPTION
VOLTAGE
TOLERANCES
CURRENT
FREQUENCY
FB000030
VIVID 7 (BT’01)
230 VAC
±10%
5 A
50-60 Hz
FC000060
VIVID 7 (BT’01)
100 - 120 VAC
±10%
10 A
50-60 Hz
FC000180
VIVID 7 PRO (BT’02)
220 - 240 VAC
±10%
5 A
50-60 Hz
FC000190
VIVID 7 PRO (BT’02)
100 - 120 VAC
±10%
10 A
50-60 Hz
FC000200
VIVID 7 (BT’02)
220 - 240 VAC
±10%
5 A
50-60 Hz
FC000210
VIVID 7 (BT’02)
100 - 120 VAC
±10%
10 A
50-60 Hz
Figure 3-3 Mains Circuit Breaker and Power Rating Label
Power Rating Label