GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
4 - 48
Section 4-3 - Functional Checks
4-3-11-4
Check Post-measurement assignment labels
5.
Use the Trackball to move the cursor to the start point
of the measurement.
The cursor points to the start point for the
measurement.
6.
Press
SELECT
to anchor the start point of the
measurement.
The start point of the measurement is marked, and a
line between the mark and the active cursor indicates
the distance.
7.
Use the trackball to move the cursor to the end point of
the measurement.
The current distance value is displayed in the
Measurement result table and is instantaneously
updated when moving the cursor.
8.
Press
SELECT
to mark the end point for the
measurement.
The measurement display color on the 2D image
changes from green to red after completion of the
measurement.
The measurement result is displayed in the
Measurement result table.
9.
Trackball to the measurement result in the
Measurement result table.
The highlighted parameter label is assigned to the
measurement.
Table 4-28 Post-measurement assignment labels
Step
Task
Expected Result(s)
10.
Trackball to the actual measurement in the
Measurement Result Window.
Press
SELECT
on the Control Panel.
A Parameter menu is displayed.
11.
Trackball through the Parameter label menu to
highlight the desired label.
Press
SELECT
on the Control Panel.
The highlighted parameter label is assigned to the
measurement.
Table 4-27 Measurement Key Checks
Step
Task
Expected Result(s)