Section 7-4 - Acquisition Diagnostics
7-4-5-2 Front End Alignment Procedure (cont’d)
Front End Alignment Description
-
Each A/D converter will get 0V input (by setting ATGC to minimum).
-
One by one channel will be digitized and the result put in the IQ buffer on the RFT board.
-
The BEP will then read the IQ buffer and computes a value per channel. These computed
values are compared to an upper/lower limit.
-
If all read values are ok, the message PASSED will appear. If values exceeding the limits are
read, the system will prompt you with FAILED and a message will tell you which BF board is
bad.
-
The resulting offsets for each channel are stored in a file on the Backend Processor’s hard
disk.
-
The values can be plotted in a diagram (click SHOW GRAPH).
6.) Always reboot the scanner after Front End Alignment is finished and verify proper operation.
NOTE:
Do Not select “Cal Reset” after performing calibration. This will destroy the file located on the
Backend Processor and the image quality will not be optimized.