GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 3 - Installation
3 - 43
3-9-3-6 Verification of a Connection Set Up to an EchoPAC PC Work Station (cont’d)
•
In the example in
Figure 3-43
, the red “X” next to the computer sign (
), means that the check
failed. The cause may be;
-
the setup of the network information, either on Vivid 7 / Vivid 7 PRO or on the EchoPAC PC, or
-
the network connection is broken somewhere between the Vivid 7 / Vivid 7 PRO and the
EchoPAC PC.
•
A green light next to the computer sign means that a network connection is in place.
Figure 3-43 Verification of connection failed
Red “X”