GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
5 - 78
Section 5-11 - Circuit Boards
5-11-12
Patient I/O
5-11-12-1
General Description
The Patient I/O contains the electronics for ECG, Phono and two analog inputs. One of the analog inputs
has programable gain. The module extracts respiration from ECG. The patient leads can be rotated by
software (i.e. it is possible to change between Lead I, II and III by pushing a button) The electronics also
contains a pacemaker detector. The four inputs are separately isolated because of safety requirements.
5-11-12-2
Location in the Unit
Inside the Backend Processor with the connector panel available from the front of the system.
5-11-12-3
Inputs
•
ECG
•
Phono
•
Two Analog Inputs
5-11-12-4
Outputs
Serial Trace Data via Internal I/O to the SPD Board